Model 7000 Series
Water Monitoring Instruments
The Series 7000 data recording instruments are assembled from standard
components to meet each customer's individual requirements.
An internal clock,
re-settable by user, tags each data reading with the
date and time at which it was taken.
The housing is our time-tested "hard coat anodized" aluminum
cylinder with patented G.O. "spaghetti line" end cap locks.
Housings are pressure rated to 5000 meters (sea water) when properly
sealed.
The following chart helps illustrate the "Electronics" and "Environmental
Sensors" which are available for each instrument. The specifications
are the same for the various instruments listed.
Pressure Temp. Tilt Serial RS232 Mem Series Name Dir S Q Conductivity Ext Int. X Y Out Cable 128/512K ____________________________________________________________________________________________________ Model 7051 DepthGauge X * * * X * * Model 7051Q Tide Gauge Quartz X * * * * X * * Model 7011U Inclinometer X * * * X X * X * * Model 7070 Thermograph * * X * X * * ____________________________________________________________________________________________________ X = Standard factory item * = Optional item S = Strain gauge pressure sensor
Q = Quartz (high accuracy) crystal pressure sensor
Specifications
Time GMT: Internal Quartz Crystal Clock, Accuracy +/- 1.0 sec/day.
User settable.
Temperature: Range: -5 C to +45 C
Accuracy: +/- 0.25 C Resolution: .050 C
Response Time: 30 seconds
External Sensor: 2 seconds
Pressure: Pressure Sensor Ranges:
0-15 PSI 34 feet 10 meters
0-30 PSI 67 feet 20 meters
0-100 PSI 224 feet 68 meters
0-500 PSI 1122 feet 341 meters
0-2000 PSI 4487 feet 1368 meters
0-5000 PSI 11218 feet 3419 meters
0-10000 PSI 22436 feet 6838 meters
Pressure housing is limited to 5000 Meters.
Accuracy: 0.5% of range Resolution: 0.1%
Accuracy (Quartz Sensor) +/- .01% Resolution (Quartz Sensor) +/- 0.005%
Sampling Rates and Intervals are configured on screen with IBM/PC/AT.
Conductivity: Toroid Inductive Coil encased in Teflon
Accuracy: +/- 1.0mS/cm
Range: 0 to 75 mS/cm
Resolution: 0.10 mS/cm
Direction: Magnetic Flux Gate
Accuracy +/- 2.0 Resolution: +/- 1.0 Range: 0 to 360
Tilt Sensor
Accuracy: +/- 0.1 Resolution: +/- 0.05 Range: 0 to 90
Pressure Case: (all instruments) 5,000 Meter (8,000 psi) anodized aluminum casing with o-ring end seals and the G.O. patented "spaghetti line" end cap locking systems.
Case Dimensions: Cylinder- 51.4 cm (20.25 in) x 10.5 cm diameter (4,125 in).
Instrument Weight: In Air- 6.9kg (15.17 lbs) maximum.
Return to the Product Catalog